Save your data if you did not use the Autosave function.
-
-
-
Lower the objective with the escape button. You can check that the Z-drive value should decrease to 499/500.
-
Never take the sample if you did not go to escape
-
Take out the frame with your sample.
-
-
-
Remove excessive oil with KIMTECH tissue
-
Take new tissue, put few drops of isopropanol and clean the objective again.
-
Switch back to 4x magnification lens
-
Go and hold the escape button on the microscope body for 2-3 seconds to reset the focus position.
-
Install back the water dispenser.
-
-
-
Click on the Discharge water
-
Take the KIMTECH tissue and blot the objective gently with a tissue to dry it.
-
Switch back to 4x magnification lens
-
Go and hold the escape button on the microscope body for 2-3 seconds to reset the focus position.
-
-
-
Close the software
-
Shutdown the computer
-
-
-
Switch off both cameras
-
Switch off the main microscope controller
-
-
-
Close the environmental box
-
If you used 1.5x Lens, always put there back the 1.0x lens for the next user!
-
Almost done!