Introduction
This guide describes how to close a session at the Zeiss Gemini 450 SEM at the Center for Microscopy and Image Analysis, University of Zurich.
-
-
Select "EHT" and set it to "EHT Off". Use the Z controller to move stage to a low position.
-
Press "Exchange" on the control panel. The stage will move the sample holder to the exchange position.
-
Wait until the transfer LED on the airlock is on and not flashing anymore. The airlock valve will open.
-
Use gloves
-
Use the white piece at the far end of the rod to slowly insert and attach it to the holder inside the microscope.
-
Screw the rod to the sample holder and retract the rod to the lock position.
-
Press "Resume" on the control panel. The airlock valve will close.
-
-
-
Press "Vent" on the Airlock panel
-
Open the airlock door
-
Unscrew the sample holder and take it out.
-
Close the airlock and press '"Store".
-
Close SmartSEM software and log off.
-