Introduction

This guide describes how to close a session at the Zeiss Gemini 450 SEM at the Center for Microscopy and Image Analysis, University of Zurich.

  1. Select "EHT" and set it to "EHT Off". Use the Z controller to move stage to a low position. Press "Exchange" on the control panel. The stage will move the sample holder to the exchange position. Wait until the transfer LED on the airlock is on and not flashing anymore. The  airlock valve will open.
    • Select "EHT" and set it to "EHT Off". Use the Z controller to move stage to a low position.

    • Press "Exchange" on the control panel. The stage will move the sample holder to the exchange position.

    • Wait until the transfer LED on the airlock is on and not flashing anymore. The airlock valve will open.

    • Use gloves

    • Use the white piece at the far end of the rod to slowly insert and attach it to the holder inside the microscope.

    • Screw the rod to the sample holder and retract the rod to the lock position.

    • Press "Resume" on the control panel. The airlock valve will close.

  2. Press "Vent" on the Airlock panel Open the airlock door Unscrew the sample holder and take it out.
    • Press "Vent" on the Airlock panel

    • Open the airlock door

    • Unscrew the sample holder and take it out.

    • Close the airlock and press '"Store".

    • Close SmartSEM software and log off.

Finish Line

One other person completed this guide.

José Maria Mateos Melero

Member since: 04/28/2019

1,131 Reputation

16 Guides authored

Team

ZMB Staff Member of ZMB Staff

11 Members

142 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 1

Past 7 Days: 5

Past 30 Days: 23

All Time: 874