Introduction
All hardware components and computer should be on. If not contact the ZMB responsible staff member.
-
-
All basic hardware components (COMET, airpump, computer) should always be on. If not, contact the responsible ZMB staff.
-
Log in with your ZMB credentials
-
Wait for Feedback form to appear and always check Device Status Information within the colored box.
-
Start COMET control software by double-clicking on the icon
-
Should any issues with the system or software occur, use the snipping tool to send a copy to the ZMB staff. Should you close the form the ?Feedback shortcut is on the desktop.
-
-
-
Begin on the HOME screen, check and address the NEXT ACTIONS listed in HOME.
-
Count down on Buffers. Do not prepare more than what is needed for your run.
-
! Red Circled Exclamation Marked actions must be performed before proceeding with reagent, protocol, or run setup.
-
Perform actions listed. Common actions: Imaging Buffer (IB) every 48hrs, Cleaning procedures as instructed.
-
Restart COMET Control Station (computer) and (software) once per week, done by user if action is listed.
-
Reservoir cleaning. Must be performed after every run (ideally within 48hrs).
-
Restart of the COMET Instrument and Fluidics cleaning is done once per month by ZMB staff. COMET is blocked in advance in the booking system on scheduled date.
-
-
-
Create new reagents
-
1. In COMET Control Software, add the necessary reagents to the reagent library: PLAN > Reagents > ADD NEW
-
Naming Convention: PI initials Lab_User initials_Antibody_Company_ Catalog#_extra info (if you use different dilutions of the same antibody or the dilutions change for different tissues include this information here). Also add the lab TAGS to your reagent for further help grouping and locating reagents.
-
Once a reagent is in a protocol you can't edit the following fields: NAME, category (secondary/primary/etc.), Channel, target species, host species, or dilutions.
-
For each new reagent, fill in at least the required fields: Name, Category, Channel (if applicable), Dilution (if applicable), Target genes (if applicable), Tails (if applicable)
-
2. For some reagent categories, a mix of reagents can be reagents can be created by combing reagents already existing in the library; to do so, select a "mix" category in the Add new page and select the reagents to mix from the drop-down menu
-
-
-
1. In COMET Control Software, create teh necessary protocols: PLAN> Protocols> ADD NEW
-
a) Select a template from the list (see section 3 of this short guide and UM for more details on the available templates). b) Chose the parameter for the selected template. c) Press OK to create a new protocol
-
2. On the protocol edition page, adjust the PROTOCOL PARAMETERS in the upper banner (valid for the whole protocol). See UM section Create and edit protocols for more details on those parameters.
-
3. On the protocol edition page, adjust the parameters in CYCLE DETAILS for each cycle. See UM section Create and edit protocols for more details on those parameters.
-
4. Click SAVE to save the newly created protocol.
-
Protocols with remaining errors (red dots) can be saved as drafts but are not executable (indicated by the red cross X). By hovering over the red dot, see the rule conflict, which prevents the protocol from being executable.
-
Once all mandatory rules are fulfilled within one protocol (only green or orange (warning dots), it can be saved and executed (indicated by a blue check).
-
-
-
1.) In COMET Control Software, create a run composed of the desired protocols by following the run creation wizard: Plan > Runs> ADD NEW
-
a) Select 1 to 4 executable protocols using the + button (s). The same protocol can be added multiple times (max 4).
-
b) Drag and drop the selected protocols to assign them to stainers 1 to 4.
-
c) Choose RUN PARAMETERS (see I< section Create and edit runs for details on those parameters) and click NEXT.
-
2.) On the SUGGESTED ALLOCATION page, see suggested allocation of reagents based on the selected protocols (highlighted reservoirs will be used for the execution of the run) and click NEXT.
-
3.) On the REVIEW page:
-
a) Review run summary. Optional: edit run name, add notes. b) Optional: generate the preparation PDF for the created run by clicking on PREPARATION PDF.
-
c) click SAVE to save the newly created run.
-
-
-
In COMET Control Software, follow the EXECUTE wizard to select protocol(s) or run, load reagents, load slide(s) and start your run.
-
1.) In the first step of the EXECUTE wizard, choose if you wish to select individual protocols or a run using the toggle.
-
In SELECT PROTOCOLS:
-
a) Select 1 to 4 executable protocols using the + button(s). The same protocol can be added multiple times (max 4).
-
b) To change the protocols assignment to stainers, drag and drop a selected protocol and assign it to a particular stainer.
-
c) Choose RUN PARAMETERS (see UM section Select run/protocols for details on those parameters)
-
In SELECT RUN: Select the run to be executed.
-
Then, click NEXT
-
-
-
2.) In the second step of the EXECUTE wizard, review the allocation of reagents.
-
Highlighted reservoirs indicate the allocation suggested by the system based on the selection of protocols.
-
Click NEXT to accept, or
-
Manually force the allocation of a reagent to a specific reservoir in INSTRUMENT > Reservoirs.
-
Then, come back to EXECUTE.
-
-
-
1.) In the third step of the EXECUTE wizard, prepare and load the reagents:
-
a) Generate the preparation PDF by clicking on PREPARATION PDF Even if already done earlier as here the Preparation PDF takes into account the current reservoirs status.
-
b) Prepare the reagents according to the preparation PDF.
-
c) Open the Reagent module door.
-
d) Load the reagents on COMET instrument as instructed by the software (see UM section Handling for details on how to handle reservoirs and bottles). For each loaded reagent, click LOAD in the COMET Control Software.
-
Important: Reservoirs and volumes must precisely match those indicated in the software. If needed, edit loaded volumes in the software.
-
e) Once all reagents have been loaded, click NEXT.
-
f) Close the Reagent module door
-
-
-
In the fourth step of the EXECUTE wizard, load your tissue slides and one COMET Chip per slide:
-
a) In COMET Control Software, review the protocol assignation to each stainer, adjust if needed.
-
b) If needed, bring the stainer to be loaded in the front position by clicking on BRING TO FRONT.
-
c) ON COMET Instrument, open the Staining and Imaging module door.
-
d) Open the lid of the front stainer. Check that the two o-rings are present and correctly positioned.
-
Remove any dust with ethanol wetted Kimwipe and by using dust blower bulb. Dust can ruin experimental autofocus finding.
-
e) Take the tissue slide out from the buffer in which it is stored and remove excess buffer on the slide using a KimTech wipe both on the back and front of the slide. Carefully dry around the tissue without damaging it. Ensure the back of the slide is entirely dry.
-
f) Insert the tissue slide in the stainer by sliding it horizontally, with the tissue facing upwards.
-
-
-
g) Place a new COMET Chip within the guiding pins, seal facing downwards. The cut corner of the chip must be placed on the top left corner of the stainer.
-
h) Close the stainer lid until the lever clicks to secure it in closed position.
-
i) Move the slide to have the region of interest right under the imaging window of the chip.
-
j) Close the Staining and Imaging module door.
-
k) In COMET Control Software, click LOAD SLIDE to start the slide loading protocol. Wait until its end, indicated by a slide icon and a lock symbol on the stainer.
-
l) On COMET instrument, open the Staining and Imaging module door, and inspect the stainer: no bubbles should be visible. In case of bubbles, refer to IM section Troubleshooting and Repair- Slide loading.
-
m) optional enter slide ID. n) Close the Staining and Imaging module door.
-
o) Repeat b) to n) for remaining slides to be loaded. p) Once all stainers have be successfully loaded, click NEXT in COMET Control Software.
-
-
-
In the fifth step of the EXECUTE wizard, review all parameters of the run. If any change is needed, click BACK to return to any previous step in the wizard.
-
Once all parameters have been reviewed and are correct, click START to begin the automated execution of the run.
-
Check that intial priming proceeds before leaving the system. Lunaphore support can help with stalled experiments if you start within business hours.
-
-
-
When a run is complete, check the pop-up in COMET Control Software to see its status. If the status is orange or red, make sure to later check the protocol report(s) to see the details of the warning(s)/error(s).
-
Close the pop-up. Preview images for completed protocols can be seen. Once the review of preview images is done, click UNLOAD.
-
In COMET Control Software, go to INSTRUMENT > Stainers to unload the slides.
-
a) Bring a stainer to the front to retrieve its slide and click OPEN STAINER. Wait until the open stainer procedure is finished (slide icon not visible anymore on that stainer).
-
b) On COMET instument, open the Staining and Imaging module door and open the lid of the stainer in the front. Remove the COMET Chip and discard it. Retrieve the slidel if the slide needs to be kept for later use and store in appropriate buffer.
-
Important: The stainer inlet/outlet o-rings may remain stuck to the COMET Chip. If the do, remove them from the chip and place them back into the o-ring grooves on the stainer. If an o-ring is lost, place a new one from the spare parts provided with the system.
-
Additionally, the COMET Chip seal may remain stuck to the slide. If it does, remove it from the slide before removing the slide from the stainer.
-
c) Dry the stainer with a tissue if any liquid remains. Close the stainer lid and the Staining and Imaging module door. d) Repeat steps a) to c) for other slides to unload.
-
-
-
In COMET Control Software, check results from individual protocols in RESULTS.
-
a) Click on the protocol of interest (by default the most recent appears at the top of the list).
-
b) Click on the "Open report" icon to open the corresponding PDF report. Refer to UM, section Protocol report errors for details on errors reported, if any.
-
c) For completed protocols, click on the "Open image" icon to open the Viewer and see the acquired images.
-
d) Click on the "Open output folder" icon to open the folder containing teh output files (report and image).
-
DATA MOVER INSTUCTIONS HERE
-
Horizon and HALO software tutorial links here.
-
-
-
Within 24 hr after the end of a run, perform a Reservoir cleaning of all small and large reservoirs used during that run.
-
a) In COMET Control Software, go to CLEANING > Reservoir cleaning
-
b) Read the Description and click NEXT.
-
c) Select the reservoirs to be cleaned by clicking on them. By default, all used reservoirs are automatically selected. Click NEXT.
-
d) Open the Reagents module door and perform the required actions displayed in COMET Control Software.
-
Used 50mL Falcon should be swapped with prefilled 50mL MilliQ Water Falcons and the large 70% Ethanol Reservoir and DIW (MilliQ) water reservoir filled to 200mL before starting the clean.
-
e) Optional: remaining reagents in small reservoirs (microtubes) can be saved before starting the cleaning procedure. To do so, go to INSTRUMENT > Reservoirs, open the rack containing the reagent you want to retrieve by clicking on the corresponding lock symbol, replace the microtubes with empty ones, and store leftover reagents for later
-
f) Close the Reagent module door and click START in CLEANING > Reservoir cleaning.
-
-
-
Once Reservoir cleaning is completed, cleaned small and large reservoirs are filled with DIW (MilliQ) no user action is needed.
-
Perform the following actions for each bottle that has been cleaned:
-
a) Discard its content (bottle content is shown black in INSTRUMENT > Reservoirs).
-
b) Wash it in the sink with soap and hot water.
-
c) Fill bottle with DIW or fresh buffer (never leave bottles empty!) and place it back on the instrument.
-
d) In INSTUMENT > Reservoirs, allocate the corresponding reagent to that bottle.
-
Important: Always keep the bottle content and volume up to date in COMET Control Software to avoid any confusion, which could lead to erroneous protocol results.
-
Users are responsible for coordinating reservoir cleaning with next user after their run. Additionally ZMB staff will block the system once a month for Fluidics cleaning to ensure the instument upkeep by cleaning all fluidic lines using dedicated cleaning solutions. Also it is not recommended to shut down the COMET.
-