Introduction
How to log off or shut down the Leica SP8 Falcon confocal laser scanning microscope located at the Irchel Campus, room Y42-H-81.
Please find detailed information about the system setup here.
-
-
Go to the "Open Projects" tab.
-
Save your data
-
using the "Save All" icon on top,
-
or click the save sign behind the "Project" if you only wish to save specific projects.
-
Please save your data on your core storage (network path: \\files.core.uzh.ch\).
-
Please follow our instructions here on how to access your data.
-
-
-
You can save your settings and reuse them by loading them in your next imaging session. By default the following settings will be loaded: laser intensities, detector settings, averaging and accumulation settings.
-
Save your sequential scan settings if applicable.
-
In your next imaging session you just need to activate "Seq" and click "Load" in order to retrieve your scan settings.
-
Please save the data on your core storage (network path: \\files.core.uzh.ch\) in the "home" folder to be able to access it after more than 4 months.
-
-
-
Lower the objectives and remove your sample.
-
Clean the immersion objective(s) you have used with the available "KIMTECH" wipes and 100% EtOH.
-
Always clean the lenses in concentric circles and from the center to the edge. Clean also the sides of the objective(s).
-
Please make sure that other microscope parts (e.g. stage) and desk are also properly clean.
-
-
-
Select the 10x dry objective via the "LAS X" software.
-
-
-
Go to "Laser Overview".
-
Set the "Frequency" to 80 MHz.
-
Go to "Configuration".
-
Select "Laser Config".
-
Switch "OFF" the lasers .
-
Check the booking system if there is another booking within the next 2 hours.
-
If YES, close the "LAS X" software and Sign-Out.
-
If NO, close the "LAS X" software, shut-down the PC and follow the next step.
-
-
-
Only follow this step if there is no booking within the next 2 hours.
-
On the right side of the table:
-
Turn the "Laser Emission" key to "OFF-0".
-
Switch OFF the "PC/Microscope", "Scanner Power" and "Laser Power" switches.
-
Switch OFF the red button (underneath the table on the left).
-
-
-
Follow this step if environmental control was applied during your experiment.
-
37°C and 5% CO2 are set as a standard. If you have changed it during your experiment, please set it back to the standard values.
-
Switch off the cube (temperature control).
-
Switch off the brick (CO2 control).
-
Close the tap on the wall dedicated to CO2 supply to the SP8 Falcon.
-
Check if CO2 environmental control is being used by the other microscope in the room (SP8 STED).
-
If NO - Close the gas bottles used.
-
If YES - Close only the N2 gas bottle exclusively dedicated to the Falcon (if it was used).
-