Store the stock solutions at room temperature (RT) or 2-8 °C if storing for longer than 3 months. Discard
the solution if it becomes cloudy or shows a large amount of precipitate. Store the stock solutions at room temperature (RT) or 2-8 °C if storing for longer than 3 months. Discard
the solution if it becomes cloudy or shows a large amount of precipitate
-
-
Fill each of the PTM tanks with 1.5 L of the appropriate buffer (H or L).
-
[Optional, use if you wish to program a preheating] In the menu “1 – Set Clock and Delay Start” set the current day and time.
-
In the menu “2 – Setup Cycle (Time and Temp)” set the “Time to run” to 01:00 h and the “Temperature” to 102 °C, or to the optimal time and temperature for your tissue and markers.
-
In the tab “3 - Preheat Settings”, set the “Preheat Temp” to 85 °C and the “Cool End Temp” to 65 °C.
-
[Optional, use if you wish to program a preheating] Set the desired “Preheat Start Time” and select the preheat mode “One Time” or “Continuous” for a single or daily preheating.
-
Place the lid on the tank, latch the PTM lid.
-
-
-
If the PTM was not preheated previously, press “Run” in the home tab to start the preheating of the tank(s)
-
Once 85 °C is reached, mount the slides into the slide rack. Open the PTM, remove the tank lid and insert the slide rack. Place lid back and latch the PTM lid.
-
Press “Run” in the home tab to start the Heating cycle.
-
When the heating cycle is finished and the tank has cooled down to 65 °C, a beep will sound, and the lid will unlock: take the rack(s) out of PTM and rinse the slide(s) with Lunaphore Multistaining Buffer.
-
Place and keep the slide(s) in fresh Multistaining Buffer at room temperature until staining.
-