-
-
1. SEM and FIB are switched on. 2. Bring sample to coincident point. 3. Center on a border of the grid. 4. FIB apertures have been initialized.
-
In the FIB Control Menu. Select Align tab.
-
Select the 30kV/50pA probe current (reference probe current)
-
Focus the specimen surface: Use Focus and Stigmation. Click on Save.
-
Change the probe current and repeat these steps for each current.
-
-
-
The beam shift correction is used to correct the different beam positions of the different probe current settings.
-
The 30kV:50pA probe current is the reference current where no correction is applied.
-
Select the crosshairs and bring a recognizable feature to the center.
-
In the FIB Control/Align tab, select the next smaller probe current from the drop-down menu.
-
If the feature is shifted to another position: Move the feature to the original position by using the beamshift correction. Click on Beam Shift Cor.
-
Move the crosshairs with the sliders.
-
Click on Save to save the adjustments. Repeat these steps for all smaller probe current.
-
Now before adjusting the higher probes switch back to the 30kV/50pA and if necessary center the feature by moving the stage before changing to the following higher currents and proceed as above for the rest of probe currents.
-
Almost done!
Finish Line