Skip to main content

v2.0

by Joana Raquel Delgado Martins

Introduction

On the MesoSPIM, the shutters can be in manual mode after initializing the system. This leads to no laser emission when imaging or to a situation where both illumination paths are emitting light at the same time. In this guide we explain you how to fix those issues.

  1. Check that the lasers are on and that the "Emission" LED is also on. Press "Live" in the MesoSPIM control software and check if the LED of the shutter is "on" in the corresponding illumination path (left or right). If the LED is "off" then the shutter is in manual mode and has to be changed to external trigger mode:
    • Check that the lasers are on and that the "Emission" LED is also on.

    • Press "Live" in the MesoSPIM control software and check if the LED of the shutter is "on" in the corresponding illumination path (left or right).

    • If the LED is "off" then the shutter is in manual mode and has to be changed to external trigger mode:

    • 1. Press once "ENABLE" to leave in “on” position

    • 2. Press and hold "ENABLE" for few seconds until the LED blinks and switch it off.

    • Done.

  2. Press live in the MesoSPIM control software and check if the LEDs of both shutters are "on" (left and right). If both LEDs are on, then the shutters are in manual mode and have to be changed to external trigger mode.
    • Press live in the MesoSPIM control software and check if the LEDs of both shutters are "on" (left and right).

    • If both LEDs are on, then the shutters are in manual mode and have to be changed to external trigger mode.

    • Press and hold "ENABLE" for few seconds until the LED blinks and switches off. Do this for both sides.

    • Done.

Finish Line

José Maria Mateos Melero

Member since: 04/28/2019

969 Reputation

12 Guides authored

Team

ZMB Staff Member of ZMB Staff

11 Members

135 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 1

Past 7 Days: 1

Past 30 Days: 5

All Time: 257