Skip to main content

v2.0

by Joana Raquel Delgado Martins

Introduction

This guide from the Center for Microscopy and Image Analysis will explain how you properly adjust the motorized correction collar of the following objectives available on ZMB instruments:

  • 93x / 1.3 GLYC objective mounted at the Leica SP8 inverse STED 3X (Irchel).
  • 63x / 1.2 WATER objective mounted at the Leica SP8 inverse (Gloriastrasse).
  • 25x/ 1.0 WATER objective available at the Leica SP8 MP DIVE Falcon (Irchel).

Precise adjustments of the motCORR can compensate for deviations in the coverslip thickness, refractive index mismatches/specimen inhomogeneities as well as temperature changes to ensure restoration of optimal resolution, signal intensity and penetration depth.

  1. Make sure you are in  the "Acquire" tab. Sample should be focused and lasers set. In the XY dialog open the "motCORR Collar Settings" dialog by clicking the arrow symbol.
    • Make sure you are in the "Acquire" tab.

    • Sample should be focused and lasers set.

    • In the XY dialog open the "motCORR Collar Settings" dialog by clicking the arrow symbol.

    • Configure your setting by

    • adjusting the slider,

    • directly entering a value into the input field, or

    • alternatively via the control panel (one needs to add the desired setting).

    • Follow the next steps for correct adjustment: Steps 2 and 3 applicable for thin samples close to the coverslip, Steps 4 and 5 applicable for any kind of sample.

  2. Only applicable if your sample is located close to the coverslip. If thick samples are used, the correction ring is best adjusted at the current z-level. Follow Step 4 in this case. Activate a laser line (e.g. 550nm). 488 nm recommended. Choose "xzy"- Scan Mode.
    • Only applicable if your sample is located close to the coverslip. If thick samples are used, the correction ring is best adjusted at the current z-level. Follow Step 4 in this case.

    • Activate a laser line (e.g. 550nm). 488 nm recommended.

    • Choose "xzy"- Scan Mode.

    • Activate a "PMT" and place the detection window underneath the chosen laser line.

    • Choose appropriate LUT. "Spectrum" recommended.

    • Open the "AOBS Configuration" window and set it to "Reflection".

  3. Click "Live"  to start live acquisition. Adjust "Gain [V]" and laser power appropriately without saturation. Adjust the correction ring to minimize reflection signal width corresponding to maximum intensity.
    • Click "Live" to start live acquisition. Adjust "Gain [V]" and laser power appropriately without saturation.

    • Adjust the correction ring to minimize reflection signal width corresponding to maximum intensity.

    • Not optimized.

    • Optimized.

    • If the result looks satisfying the adjustment is done.

  4. Applicable for cells on cover glasses as well as thick specimen like tissue samples. Best to work with the HyD detector set to counting mode.
    • Applicable for cells on cover glasses as well as thick specimen like tissue samples.

    • Best to work with the HyD detector set to counting mode.

    • Focus your sample in the fluorescence mode. Make sure you are in the "xyz"- Scan Mode.

    • Activate "Auto-scaling" by clicking the "M" (when activated it displays an "A").

    • Carefully adjust the correction collar in one direction. The focus has to be constantly corrected.

    • Check whether the intensity and contrast of your specimen are improved - can be easily observed by the photon counts.

    • If yes, continue in the same direction and constantly check your image.

    • Finally, do this iteratively in both directions and adjust the ring to the maximum intensity and highest contrast.

  5. Same procedure can also be applied to the profile view of your sample. Choose "xzy" scan mode.
    • Same procedure can also be applied to the profile view of your sample.

    • Choose "xzy" scan mode.

    • A dotted line appears in the image. Choose the area where you would like to see your cross section (xz view).

    • If the correction ring is set poorly your structures show low intensity, and have low resolution (large point spread function (PSF)).

    • Adjust the correction ring to minimize the PSF size corresponding to maximum intensity.

    • If the correction ring is adjusted optimally the intensity should have increased (can be observed by the photon counts) and the elongation of the PSF decreased.

Finish Line

Jana Döhner

Member since: 04/10/2019

1,393 Reputation

20 Guides authored

Team

ZMB Staff Member of ZMB Staff

11 Members

135 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 0

Past 7 Days: 8

Past 30 Days: 29

All Time: 691