Introduction
How to shut down the Leica SP8 upright confocal laser scanning microscope located at Irchel Campus, in the Medical Virology (Y36-M-92).
Please find detailed information about the system setup here.
-
-
Go to the "Open projects" tab.
-
Either use the "Save All" icon on top, or click the save sign behind the "Project".
-
Choose your personal zmb data path folder on files.core.uzh.ch.
-
Please follow our instructions here on how to access your data.
-
-
-
Lower the stage and remove your sample.
-
Clean the immersion objective(s) you have used with the available "KIMTECH" wipes and 100% EtOH.
-
-
-
Go to Configuration -> "Laser Config". Check in the booking system if there is another booking shortly after you:
-
If "YES":
-
keep the lasers on standby,
-
close the "LAS X" software and Sign out of your Windows profile.
-
If "NO" :
-
switch OFF all lasers ,
-
close the "LAS X" software,
-
shut-down the PC and follow the next step.
-
-
-
Switch OFF the fluorescence lamp.
-
Turn the "Laser Emission" key to "Off-1" (control unit underneath the table) and switch OFF the "Scanner Power", "Laser Power".
-
Switch OFF the power knob (on the PC table).
-