Introduction
In this guide of the Center for Microscopy and Image Analysis we show how to shut down the Leica SP8 inverse confocal laser scanning microscope located at Y42-H-81.
-
-
Go to "Open Projects" tab.
-
Save your data
-
by either right-clicking on the "Project",
-
using the "Save All" icon on top,
-
or clicking the save sign behind the "Project".
-
Please save your data on your core storage (network path: "\\files.core.uzh.ch\").
-
Please follow our instructions here on how to access your data.
-
-
-
You can save your settings and reuse them by loading them in your next imaging session. By default the following settings will be loaded: laser intensities, detector settings, averaging and accumulation settings.
-
Save your sequential scan settings if applicable.
-
In your next imaging session you just need to activate "Seq" and click "Load" in order to retrieve your scan settings.
-
To save only single channel settings:
-
Click the "Save-Sign" and choose an appropriate name for saving.
-
Click "Ok".
-
You can now find your single channel setting under "User Settings".
-
-
-
Lower the objectives and remove your sample.
-
Clean the immersion objective(s) you have used with the available "KIMTECH" wipes and 100% EtOH.
-
Always clean the lenses in concentric circles and from the center to the edge. Clean also the sides of the objective(s).
-
Please also take care that immersion media has been cleaned off of all the other microscope parts which got in touch with.
-
-
-
Switch to the 10x dry objective.
-
-
-
Only follow this step if environmental control was applied during your experiment.
-
37°C and 5% CO2 are set as a standard. If you have changed it during your experiment, please set it back to the standard values.
-
Switch off the heating cube (underneath microscope table).
-
Switch off the brick (CO2 control).
-
Close the dedicated valves of the gas supplies.
-
-
-
Go to "Configuration" and choose "Laser Config".
-
If applicable, lower the power to 0% for the STED lasers.
-
Switch OFF all lasers.
-
If STED was used, turn OFF the Laser Emission Keys of the 592nm and 660nm STED lasers.
-
Check in the booking system if there is another booking within the next 2 hours.
-
If YES, close the "LAS X" software and Sign-Out.
-
If NO, close the "LAS X" software, shut-down the PC and follow the next step.
-
-
-
Only follow this step if there is no booking within the next 2 hours.
-
Switch-OFF the fluorescence lamp.
-
On the main switch board
-
turn the "Laser Emission" key to "Off-0",
-
and switch OFF the "Scanner Power", "Laser Power" and "PC/Microscope".
-
Take care the PC has been fully shut-down before.
-
Cancel: I did not complete this guide.
One other person completed this guide.