Introduction
How to log off or shut down the Leica SP8 MP DIVE FALCON microscope located at the Irchel Campus, room Y42-H-77.
Please find detailed information about the system setup here.
-
-
Go to the "Open Projects" tab.
-
Save your data
-
using the "Save All" icon on top,
-
or click the save sign behind the "Project" if you only wish to save specific projects.
-
Please save your data on your core storage (network path: \\files.core.uzh.ch\).
-
You can find how to access your data here on our webpage.
-
-
-
Move the objective upwards (away from your sample).
-
Clean the immersion objective(s) you have used with the available "KIMTECH" wipes.
-
Please make sure that other microscope parts (e.g. stage) and desk are also properly clean.
-
-
-
Go to "Laser Overview".
-
Set the "Frequency" of the WLL to 80 MHz.
-
Go to "Configuration".
-
Select "Laser Config".
-
Switch "OFF" both the WLL and the MP2 laser .
-
Check the booking system if there is another booking within the next 2 hours.
-
If YES, close the "LAS X" software and Sign-Out.
-
If NO, close the "LAS X" software, shut-down the PC and follow the next step.
-
-
-
Only follow this step if there is no booking within the next 2 hours.
-
On the right side of the table:
-
Turn the "Laser Emission" key to "OFF-0".
-
Switch OFF the "PC/Microscope", "Scanner Power" and "Laser Power" switches.
-
Switch OFF the fluorescence lamp.
-
Switch OFF the red button.
-